Model Number N/A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.(b)(6).
|
|
Event Description
|
It was reported that while using the pulsavac wound lavage, in tkr procedure, the tip of the pulsavac came off.No adverse events were reported as a result of this malfunction.
|
|
Event Description
|
There is no additional information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.The root cause of the reported issue is attributed to the locking ring being at the low end of specification.The event cannot be confirmed.
|
|