BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1008K
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 09/27/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that one unit of an english-chinese infusor lv (large volume) had the ¿administration tube set cracked¿.This was identified prior to use on a patient.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d10, h3, h4 and h6: h4: the device was manufactured from september 3, 2019 - september 4, 2019.H10: the actual device was received for evaluation filled with fluid.A visual inspection performed using the naked eye found broken tubing at the junction of the white flow restrictor.A portion of the broken tubing remained inside the solvent-bonded area of the flow restrictor.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10730781
MDR Text Key212968265
Report Number1416980-2020-06608
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412022512
UDI-Public(01)00085412022512
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C1008K
Device Lot Number19J021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Date Manufacturer Received11/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-