Model Number 3100A |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 08/09/2020 |
Event Type
Death
|
Manufacturer Narrative
|
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.However, the customer checked the unit and found that the capacitor is burned and brown wire of power cord is not firmly attach to the terminal block and came-out easily whereas to blue and yellow wires which are found securely connected.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer stated that the patient is on a 3100a ventilator for 48 hours and worked fine, but unfortunately due to patient pre-exisiting bad health he was unable to survive and died.Then, the unit was disconnected from the patient and switched off.The customer confirmed that there is no allegation to the ventilator for the patient's death.
|
|