• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SITE-RITE 8 ULTRASOUND SYSTEM CONSOLE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SITE-RITE 8 ULTRASOUND SYSTEM CONSOLE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770600Y
Device Problem No Display/Image (1183)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per picc rn - unit was fine for the midline she inserted but when she got the unit back out into the hall the screen had gone black except for the bard logo in the top corner.I had her unplug all of the usb peripherals and power, switched off the battery then turned it back on and plugged in power.The unit booted up at that point fine.We re-inserted the usb peripherals and it is working again.
 
Manufacturer Narrative
H11: section a through f the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of display failure was unconfirmed.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.H3 other text : evaluation findings are in section h.11.
 
Event Description
Per picc rn unit was fine for the midline she inserted but when she got the unit back out into the hall the screen had gone black except for the bard logo in the top corner.I had her unplug all of the usb peripherals and power, switched off the battery then turned it back on and plugged in power.The unit booted up at that point fine.We re-inserted the usb peripherals and it is working again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND SYSTEM CONSOLE (REMANUFACTURED)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
MDR Report Key10731187
MDR Text Key214560679
Report Number3006260740-2020-20234
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138287
UDI-Public(01)00801741138287
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770600Y
Device Catalogue Number9770600Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Date Manufacturer Received03/09/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-