It was reported that patient presented to the emergency room on (b)(6) 2020 after receiving an appropriate shock for a ventricular fibrillation episode.Device interrogation revealed that the patient's right ventricular lead was exhibiting noise over-sensing immediately following the ventricular fibrillation termination.Decreased r-wave sensing amplitudes were also noted.Programming changes were attempted, but the noise could not be avoided.The noise was consistent with the patient's left ventricular assist device (lvad) that was implanted in 2019.Further device interrogation on (b)(6) 2020 revealed that the lead experienced a significant pacing impedance drop at the time of the lvad implant, and that the lead had its capture threshold increase.Lead was capped and replaced on (b)(6) 2020.Patient had no symptoms as a result of the lead malfunction and was stable after the procedure.
|