This will be filed to report lock line break.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with grade 4+.The clip delivery system (cds) was advanced but when the delivery catheter black fastener was tightened, the physician commented that excessive force had to be applied to tighten the fastener.Then, when the clip was unlocked, the clip arms had resistance opening.Therefore, the physician decided to not use the cds and replace it with a new cds.The new cds completed the procedure successfully.One clip implanted, reducing mr to 1+.There was no adverse patient effect and no clinically significant delay in the procedure.Based on the device analysis, the lock line was noted to be broken.No additional information was provided.
|
All available information was investigated and the reported difficult to tighten delivery catheter (dc) fastener could not be confirmed, and the reported clip difficult to open could not be tested via returned device analysis.Analysis additionally identified a broken lock line and scratches on the dc handle coupler and the actuator coupler.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history no similar complaints reported from this lot.Based on the information provided a conclusive cause for the reported difficult to tighten dc fastener cannot be determined.A conclusive cause for the reported difficult to open clip cannot be determined.The observed scratches on the dc handle coupler are likely the result of advancing and retracting the clip delivery system during use.The observed actuator coupler scratches are likely the result of maneuvering the device under tension during use.Based on the information provided the observed lock line break is likely the result of troubleshooting efforts.There is no indication of a product issue with respect to manufacture, design, or labeling.
|