Model Number N/A |
Device Problem
Failure to Cut (2587)
|
Patient Problem
Skin Tears (2516)
|
Event Type
Injury
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that during surgery the dermatome tore the skin off.This required a new graft to be taken as well as additional sutures.No additional consequences have been reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the control bar was out of position and the device was out of calibration; however, the repair was not completed because the customer is waiting on a final response and new device quote.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|
Event Description
|
No additional event information available.
|
|