Model Number A710962
Device Problems Reflux within Device (1522); Improper or Incorrect Procedure or Method (2017); Suction Problem (2170); Obstruction of Flow (2423)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/02/2020
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history record will not be performed.The device has been returned to the manufacturer and the evaluation is still pending.The investigation of the reported event is currently underway.
 
Event Description
It was reported that sometime 3 years port placement, the device allegedly malfunctioned.It was further reported that there was an initial flash back of blood.Reportedly, when the port was flushed there was a distinct bulge in the internal jugular vein.Current status of patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: one powerflow apheresis iv implantable port attached to a catheter was returned for evaluation.Visual, microscopic visual and functional evaluation were performed on the returned device.The investigation is confirmed for suction problem as the device was unable to be aspirated during functional evaluation.However, the investigation is inconclusive for the reported obstruction of flow, improper or incorrect procedure or method and reflux within device issues as the exact circumstances at the time of the reported event are unknown, and the reported event could not be reproduced in the lab.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that approximately three years port placement, the device allegedly malfunctioned.It was further reported that there was an initial flash back of blood.Reportedly, when the port was flushed there was a distinct bulge in the internal jugular vein.The current status of patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLOW APHERESIS IV IMPLANTABLE PORT, CHRONOFLEX, 9.6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key10745479
MDR Text Key213392396
Report Number3006260740-2020-20264
Device Sequence Number1
Product Code PTD
UDI-Device Identifier00801741129438
UDI-Public(01)00801741129438
Combination Product (y/n)N
PMA/PMN Number
K163001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA710962
Device Catalogue NumberA710962
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2020
Date Manufacturer Received04/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age50 YR
-
-