MAUDE Adverse Event Report: ZIMMER BIOMET, INC. BASE PLATE UNCEMENTED; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. BASE PLATE UNCEMENTED; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Information (3190)
Event Date 05/01/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: poly liner plus diameter cat: 00434903600 lot: 62433842.Glenosphere cat: 00434903611 lot: 62368151.Foreign report source: (b)(6).Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2020 - 03015.Reported event was confirmed by review of radiographs and medical records.Radiographs identified glenosphere dissociation without dislocation.Medical records noted signs of osteolysis and scapular notching.Review of the device history records identified no related deviations or anomalies during manufacturing.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient underwent a revision due to glenosphere disassociation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE PLATE UNCEMENTED
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key10756021
MDR Text Key213693232
Report Number0001822565-2020-03651
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K121543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model NumberN/A
Device Catalogue Number00434903811
Device Lot Number62562700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/22/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/23/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight83
-
-