• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500 BLOOD GAS ANALYZER; RP 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500 BLOOD GAS ANALYZER; RP 500 Back to Search Results
Model Number 10492730
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2020
Event Type  malfunction  
Manufacturer Narrative
The customer states that both patients have the same last name and confirms that blood was drawn and collected from the intended, correct patient.Customer states that he deleted the sample with the incorrect acc# under patient history and invalidated the sample.He created a new order/acc# under intended correct patient which has been edited in the blood gas instrument under recall patient results.The customer states the issue was due to user error, the issue has been resolved and the instrument is operational and wants to close the investigation.
 
Event Description
The customer states that an order was created under the wrong patient which delivered blood gas results into the wrong patient's chart in lis.Customer states that both patients have the same last name and confirms that blood was drawn and collected from the intended, correct patient.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
6107052212
MDR Report Key10756320
MDR Text Key213947258
Report Number3002637618-2020-00059
Device Sequence Number1
Product Code CHL
UDI-Device Identifier00630414589169
UDI-Public00630414589169
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10492730
Device Catalogue Number10492730
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-