It was reported that while flushing the 6-8x40mm acculink self-expanding stent system, the white tip separated from the distal end of the shaft.There was no patient involvement.A new acculink was used to successfully complete the procedure.There were no adverse patient effects, and no clinically significant delay in the procedure.No additional information was provided.
|
Visual analysis was performed on the returned device.The reported tip separation was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation determined that the reported tip separation was related to circumstances of the procedure.It is likely that during removal of the stylet, the tip was inadvertently grasped causing the tip to separate.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
|