Model Number 014407
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Skin Inflammation (2443)
Event Date 07/17/2020
Event Type  Injury  
Event Description
It was reported that, after a procedure with foot strap distractors, the patient developed skin inflammation.Patient's outcome is unknown.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.Insufficient product identification information was provided, and thus, a manufacturing record review could not be conducted.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Per subsequent e-mail, no relevant clinical information not the device will return for inclusion in this medical investigation.Therefore, a thorough medical investigation cannot be rendered.Per the complaint details, the patient¿s outcome is unknown.Therefore, no further clinical/medical assessment is warranted at this time.The complaint was not confirmed.Factors that could have contributed to the reported event include improper placement/adjustment of the device or an adverse biological patient reaction to the device materials.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOT STRAP ANKLE DSTRCTR STRL(6)
Type of Device
COMPONENT, TRACTION, NON-INVASIVE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10759818
MDR Text Key213843288
Report Number1219602-2020-01703
Device Sequence Number1
Product Code KQZ
UDI-Device Identifier03596010095329
UDI-Public03596010095329
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number014407
Device Catalogue Number014407
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-