Model Number 2545-00-940 |
Device Problem
Component Missing (2306)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
During routine instrument checks yesterday, the mdrd technician found 4 items damaged in attune sets.Each of the impactor heads were cracked.Both the femoral and rp tibial impactor heads had vertical cracks on the back; while the fb tibial impactor had a vertical crack on the front.The other item that was damaged was a size 7/8 - 10mm shim; it was missing the metal rings for the 2 holes.The mdrd technician predicts that they went missing during a wash cycle - they were not found.None of the damage produce fragments or had a delay on surgery.All 4 items were taken out of the sets and are available for return.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: examination of the returned device confirms the reported event.The need for corrective action is not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|