Model Number 364501
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2020
Event Type  malfunction  
Manufacturer Narrative
Combination product: yes.
 
Event Description
The orsiro drug-eluting stent system was chosen for the treatment.Prior crossing the target lesion a defect at the distal part of the device was noticed.
 
Manufacturer Narrative
Combination product: yes.The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed that the hypotube has fractured about 17 cm distal to the kink protector.Kinks were observed close to the fracture sites of both fragments.The crosssections of the hypotube are no longer circular but compressed and the polymer coating is plastically deformed.The findings indicate that the hypotube most probably fractured as a result of significant bending outside of the patients body.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause for the complaint event is most likely related to the handling of the device during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.75/22
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10775329
MDR Text Key214303656
Report Number1028232-2020-04749
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
P170030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/07/2021
Device Model Number364501
Device Catalogue NumberSEE MODEL NO.
Device Lot Number09195218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2020
Date Manufacturer Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-