PRISMATIK DENTALCRAFT, INC. COMFORT HARD-SOFT SPLINT; THERMOFORM MOUTHGUARD Back to Search Results
Lot Number 1107082691
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Type  Injury  
Manufacturer Narrative
The device has been returned.Once the investigation is complete a supplemental report will be submitted.Is not applicable with the exception of lot number as the device is manufactured by prescription.Is not applicable as the device is manufactured by prescription not implantable.
 
Event Description
It was reported that the patient had an allergic reaction to the comfort hard-soft splint.
 
Manufacturer Narrative
The device investigation has been completed and the results are as follows: dhr results: no dhr was available for review since the device was fabricated per physician's prescription only.Supplier (erkodent) reviewed the associated material lot and confirmed no manufacturing deviation or abnormality.Lot# e-pro 4.0-11378 (erkoloc-pro) was manufactured from 10/28/19 and was assigned with a 3 year expiration.Stock product reviewed results; no stock product was available for review since the device was fabricated per physician's prescription only.Investigation methods/results: an upper tray was returned but not in the original case.Device results are summarized below:roughness - the flange was smooth.Occlusal/lingual surface at the central incisors appeared to have been adjusted and was slightly rough.Crack - no major crack was found.Delamination - layers were intact and did not separated.Discoloration - no discoloration was found.General cleanliness - the returned device appeared to be clean, free of debris or foreign particles.The returned device was visually inspected and no defect and abnormality was observed.There was no evidence found to indicate that the reported issue was caused by the device itself.Root cause: a root cause for this complaint cannot be explicitly determined.Ifu 9091 rev 3.0 (comfort h/s bite splint instruction for use) states to use only clear, cool water to wash the device.Ifu provides warning "do not clean or soak in mouthwash; do not use denture cleanser, hot water, alcohol, hydrogen peroxide; do not place in direct sunlight".It is possible that reactions could be caused by mouthwash, toothpaste, or soaking material.However, the customer did not provide the information regarding how the patient handled and maintained the device.Glidewell research team and namsa conducted a series of testing on a similar thermoformed sleep device following iso 10993 (biological evaluation of medical devices) and the device was evaluated for potential cytotoxicity, skin irritation, delayed dermal contact sensitization and oral mucosal irritation.The test article was thermoformed with layers of erkodent material (erkoloc-pro and erkodur).The test results were listed below and summarized in biocompatibility report for sleep device (rpt 9733 rev 1.0).For cytotoxicity testing, the test article extract showed no evidence of causing cell lysis or toxicity.For skin irritation, there was no erythema and no edema observed on the skin of the animals treated with the test article.For sensitization testing, the test article extracts showed no evidence of causing delayed dermal contact sensitization.The test article showed nonirritant to the oral mucosa as compared to the control article.The device materials have been found to be biocompatible through the testing.There was no cytotoxic, sensitization, skin irritation, or oral mucosal irritation found in any of the test articles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMFORT HARD-SOFT SPLINT
Type of Device
THERMOFORM MOUTHGUARD
Manufacturer (Section D)
PRISMATIK DENTALCRAFT, INC.
2212 dupont drive
suite p
irvine CA 92612
MDR Report Key10777307
MDR Text Key214374658
Report Number3011649314-2020-00706
Device Sequence Number1
Product Code MQC
Combination Product (y/n)N
PMA/PMN Number
K121365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Lot Number1107082691
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2020
Date Manufacturer Received06/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-