• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY TIMESH STRAIGHT PLATE 1H X 1H 8 MM; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY TIMESH STRAIGHT PLATE 1H X 1H 8 MM; PLATE, BONE Back to Search Results
Model Number 015-042
Device Problem Break (1069)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the surgery, the screw slipped, the screw broke, and the connecting plate and titanium cap hole broke.This resulted in a procedure delay of 2 minutes.The procedure was completed with backup products.The patient's status at the time of the report was alive-no injury.Additional information received reported that this problem had been occurring since (b)(6) 2018; however, this was the first time the hospital had notified the distributor of the quality problems.It was noted that the titanium products need to be disinfected in the hospital supply room after unpacking, and then sent to the operating room.Similar products on the market may cause the product to break due to manual operation and other reasons during the disinfection process.This compensation mainly belonged to the obvious broken products picked out by the staff in the supply room during the disinfection or after the doctor opened the disinfection package in the operating room.
 
Event Description
Additional information received reported that a disinfecting agent was not used as part of the disinfection process.The disinfecting process that was followed was clean water shower plus high temperature and high pressure disinfection.The process of putting loose screws into the box was operated by the workers responsible for the disinfection in the hospital.It was noted that the temperature was 134 degrees celsius, and the pressure was 3bar.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIMESH STRAIGHT PLATE 1H X 1H 8 MM
Type of Device
PLATE, BONE
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
MDR Report Key10781533
MDR Text Key214801599
Report Number2021898-2020-00296
Device Sequence Number1
Product Code JEY
UDI-Device Identifier00643169620278
UDI-Public00643169620278
Combination Product (y/n)N
PMA/PMN Number
K973145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number015-042
Device Catalogue Number015-042
Device Lot NumberT200549
Was Device Available for Evaluation? No
Date Manufacturer Received12/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-