Model Number CV-180 |
Device Problems
No Display/Image (1183); Poor Quality Image (1408)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
During preparation for use, the endoscopic image was not displayed.There was no report of patient injury associated with this event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The device was returned to the service department of olympus medical systems india private limited (omsi).The evaluation of the device confirmed the followings; -distortion of endoscopic image due to the defect of the circuit board was noted.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Omsc guessed that the abnormal endoscopic image was caused by the defect of the circuit board due to aging.If additional information becomes available, this report will be supplemented.
|
|