Model Number CV-180
Device Problems No Display/Image (1183); Poor Quality Image (1408)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/06/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
During preparation for use, the endoscopic image was not displayed.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The device was returned to the service department of olympus medical systems india private limited (omsi).The evaluation of the device confirmed the followings; -distortion of endoscopic image due to the defect of the circuit board was noted.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Omsc guessed that the abnormal endoscopic image was caused by the defect of the circuit board due to aging.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10783621
MDR Text Key214641039
Report Number8010047-2020-08504
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-