(b)(6).The axium prime coil (model: apb-2-4-3d-es lot: a771108) was returned for analysis.The dispenser coil and black rubber guidewire clip were not returned for analysis.No irregularities were found with the returned packaging material.The opened outer carton and inner pouch has all seals accounted for.No damages or irregularities were found with the introducer sheath.The axium prime pusher was found broken at the break indicator.The release wire was fully retracted out of the distal pusher segment.The release wire was still attached to the proximal segment.The pusher was found kinked at 119.6cm from the proximal end of the distal segment.What appears to be dried blood was found coagulated on the shield coil.The implant coil was already detached.The implant coil was found stretched with the polypropylene filament broken.No other damages or anomalies were observed.Based on the device analysis, reported information, and customer device image, the customer¿s report of ¿pusher bent¿ and ¿prod damaged/deformed out of pkg¿ were confirmed as the pusher was found to kinked; however, the cause could not be determined.Additional damages to the device were found beyond those the customer reported.The pusher was found kinked/broken, the release wire was fully retracted out of the pusher and the implant coil was found stretched with the polypropylene filament broken.In addition, what appears to be coagulated blood was found on shield coil, indicative that the device was potentially used.As the dispenser coil and guidewire clip was not returned for analysis, any contribution of the dispenser coil and guidewire clip to the ¿pusher kink/broke¿ and ¿prod damaged/deformed out of pkg¿ could not be assessed.The investigation determined that this event was similar to an event that had already been investigated, and another investigation is not necessary.Based on the formal investigation, it is possible the damage occurred during production or upon opening the package and attempting to remove the product or after removing it from the package.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Therefore, the root cause could not be determined.If information is provided in the future, a supplemental report will be issued.
|
It was reported that when the axium coil was taken out of the packaging, it was bent on the proximal segment of the delivery wire.The physician decided to use another coil of the same size from another manufacturer to complete the procedure.There was no patient involvement indicated with this event.The physician noted that there no damage or evidence of tampering to device packaging, the proximal end of the pushwire was secured in the guidewire clip when the package was opened.
|