Model Number 71992-01 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Information (3190)
|
Event Date 08/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.The sensor is designed to report readings of 40 mg/dl to 500 mg/dl.It should be noted that this sensor does not give numeric value for readings less than 40 mg/dl.A "lo" display message indicates a reading less than 40 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
Customer reported receiving a low reading from their freestyle libre 2 sensor.Customer reported no trend arrows at the time of the call.Customer reported a low reading of "lo" (below 40 mg/dl) which was lower than lab reading of 200 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(4) has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Inspected the plug assembly, no issues were observed.The current was applied to the sensor to perform linearity testing while in the test fixture.All results were within specification.No malfunction or product deficiency was identified.
|
|
Event Description
|
Customer reported receiving a low reading from their freestyle libre 2 sensor.Customer reported no trend arrows at the time of the call.Customer reported a low reading of "lo" (below 40 mg/dl) which was lower than lab reading of 200 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
|