• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD ENDO CLIP III; CLIP, IMPLANTABLE Back to Search Results
Model Number 176630
Device Problem Break (1069)
Patient Problems No Information (3190); Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, intra-operatively, the device broke.
 
Manufacturer Narrative
The incident device has been received and is under evaluation.When the device evaluation is complete, a follow-up report will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection found that the instrument was partially applied.The clip counter was no longer active.No visual abnormalities were observed with the instrument.Functional testing noted that the instrument was applied to appropriate test media and was cycled without binding.Fourteen clips advanced into the jaws, formed properly, and were held securely in place after full formation was achieved and the firing handle was released.The next clip would not load into the jaws when the handle was cycled and the clip was moving back and forth with the pusher bar.It was reported that the device broke.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key10799727
MDR Text Key215016908
Report Number9612501-2020-01676
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057852
UDI-Public10884521057852
Combination Product (y/n)N
PMA/PMN Number
K100242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model Number176630
Device Catalogue Number176630
Device Lot NumberJ0C1442Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2020
Date Manufacturer Received03/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-