Model Number SUPPORT ARM 177
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The customer did not request service, and replaced the broken support arm by themselves.No pictures or parts were received for further investigation the support arm is a casting, and it most probably developed a crack at an earlier occasion either by overloading, or impact, and this led to the reported breaking.Previous investigations led to a redesign and a change of the manufacturing process in order to obtain a higher mechanical strength of the support arm.This change was implemented in production during september 2009.However, since no pictures, or parts provided for our investigation, the root cause of the broken support can not be established.
 
Event Description
It was reported that the support arm of the ventilator was damaged.There was no patient harm.Manufacturer ref.#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPPORT ARM 177
Type of Device
SUPPORT, ARM
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
roentgenvagen 2
solna 
MDR Report Key10800219
MDR Text Key215049177
Report Number8010042-2020-00919
Device Sequence Number1
Product Code IOY
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSUPPORT ARM 177
Device Catalogue Number6481720
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/12/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/05/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N