Model Number SPL-T
Device Problem Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device was received and evaluated.Evaluation determined that the cable was found separating from the transducer housing and frayed wires were observed sticking out due to damaged cable.The device is non-repairable.Based on evaluation findings the reported issue was confirmed.Probable cause could be due device normal wear and tear and or handling issue.
 
Event Description
It was reported that the device cord was found separating from the transducer.The generator is not turning on.No further details provided regarding the event.There was no patient involvement reported.No user injury reported.
 
Manufacturer Narrative
This supplemental report is being submitted to provide review of the device history records (dhr), unique device identifier (udi) number and investigation conclusion.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The reported failure was most likely due to user mishandling.As stated on the ifu (instruction for use) the user manual states: this product is a precision device; handle it with care.Avoid rough or violent handling, which may cause equipment damage, take special care to ensure all cables are undamaged.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI INC
136 turnpike road
southborough PA 01772
MDR Report Key10804614
MDR Text Key219550508
Report Number3011050570-2020-00139
Device Sequence Number1
Product Code FEO
Combination Product (y/n)N
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Was the Report Sent to FDA? No
Date Manufacturer Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-