• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. DISPOSABLE ELECTROSURGICAL SNARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. DISPOSABLE ELECTROSURGICAL SNARE Back to Search Results
Model Number SD-210U-15
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/19/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.There was no abnormality in the connection between the tube and the operation part.Something like tissue had adhered to the inside of the tube.Foreign material was attached to the snare loop.It is probable that the liquid leakage from the operation part was caused by the liquid entering from the tip of the tube to the operation part.It is possible that the body fluid has entered into the tube due to the protrusion and retraction of the snare loop in the body fluid.However, the cause could not be identified.Since this product is not watertight in the operation part, if body fluid enters inside the tube, it may leak from the operation part.
 
Event Description
We received the following report.*during an endoscopic mucosal resection, the subject device was used.Mucus leaked from the handle of the device.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE ELECTROSURGICAL SNARE
Type of Device
DISPOSABLE ELECTROSURGICAL SNARE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10806544
MDR Text Key225941383
Report Number8010047-2020-08671
Device Sequence Number1
Product Code FDI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K902735
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberSD-210U-15
Device Lot Number01K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-