Catalog Number 031-28J |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
|
Event Description
|
The complaint is reported as: "when the user connected the nebulizer adaptor and the flowmeter, and attached the heater and the aquapack bottle, the connecting part between the nebulizer adaptor and the flowmeter got detached.Therefore, a new unit was used instead.No harm to the patient occurred.".
|
|
Manufacturer Narrative
|
(b)(4) the sample was returned for evaluation.A visual exam was performed and it was observed that there was damage on the inner section of the thread of the adaptor.No other issues were found.Oxygen entrainment testing was performed and during the setup it was observed that the assembly of the nut adaptor and the upper body was unstable.Even with that condition, the sample was able to be tested on oxygen entrainment testing but the testing failed due to the unstable condition.After the testing finished, the adaptor was carefully disassembled from the upper body and it was visually inspected.During the visual inspection wear was found on the internal tabs.Based on the investigation performed, the complaint is confirmed.Although the complaint is confirmed, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor was most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.
|
|
Event Description
|
The complaint is reported as: "when the user connected the nebulizer adaptor and the flowmeter, and attached the heater and the aquapack bottle, the connecting part between the nebulizer adaptor and the flowmeter got detached.Therefore, a new unit was used instead.No harm to the patient occurred.".
|
|