Model Number AIA-360 |
Device Problem
Difficult to Flush (1251)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.A field service engineer (fse) was at the customer's site to address reported event.Fse confirmed reported error by reviewing the analyzer printouts, and reproduced error by attempting a startup.Fse resolved the complaint by replacing the bf wash probe.Fse successfully initialized the analyzer without error, performed quality control run without error and results were within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-360 operator's manual under chapter 7, section 7.1: list of error messages states the following: [2015] bf probe purge failure is generated when purging by the bf probe is abnormal.The operator is instructed to clean up the wash probe tip or replace it.Contact the service department.The most probable cause of the reported event is due to failure of the bf wash probe.
|
|
Event Description
|
A customer reported getting error message 2015 "bf probe purge failure" during startup of the aia-360 analyzer.Technical support specialist (tss) instructed the customer to clean the electrodes with alcohol and repeat the installation wizard to prime, but error persisted.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for estradiol (e2).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
The bf wash probe was returned to tosoh instrument service center for investigation.Functional testing of the probe could not confirm reported event.The most probable cause of the reported event could not be duplicated.
|
|
Search Alerts/Recalls
|