SMITH & NEPHEW, INC. JOURNEY DCF AP FEM CUT BLK 5; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 74012415 |
Device Problem
Break (1069)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/21/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during tka procedure the journey dcf ap fem cut blk 5 broke.This happened during use outside the patient.There was no delay.Procedure concluded with a s&n back up device.No injuries or other complications were reported at this time.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the posts are broken off the journey dcf ap fem cut blk 5.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|