• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Model Number ER420
Device Problems Failure to Fire (2610); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/30/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: u93z6p.Investigation summary: the analysis results found that the er420 device was returned with no apparent damage, with a clip in the jaws.Three open clips and three conforming clips were also returned inside of a plastic bag.The device was tested for functionality and during the analysis the device was cycled and it fed, retained, and formed the remaining eleven clips as intended.The event described could not be confirmed as the device performed without any difficulties noted.Although there is no direct evidence of use error, the instructions for use do contain the following: "caution: fully squeeze the trigger and then release to load the first clip into the instrument jaws.Failure to completely squeeze the trigger can result in clip mis-loading.Position the jaws with the clip completely around the vessel to be ligated.Fully squeeze the trigger on each firing.Do so by pulling back on the trigger even after a sharp ¿click¿ is heard and until you touch plastic trigger to plastic handle (plastic to plastic).Fully release the trigger after firing.A second ¿click¿ should be heard indicating the instrument is ready for the next firing.The next clip is automatically advanced as the trigger is released.Inspect the instrument jaw tips after each use to ensure a new clip is present before the next firing." it should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that during a laparoscopic cholecystectomy, the surgeon felt the handle could not be grasped fully during use.The two clips were fed at the same time.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key10840206
MDR Text Key216410962
Report Number3005075853-2020-06006
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012597
UDI-Public10705036012597
Combination Product (y/n)N
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberER420
Device Catalogue NumberER420
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2020
Date Manufacturer Received10/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-