• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE CONN./EXT. DL; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE CONN./EXT. DL; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Model Number 050-87216
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Inflammation (1932); Peritonitis (2252)
Event Date 01/01/2018
Event Type  Injury  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.  clinical investigation: a temporal relationship does not exist between ccpd therapy utilizing the liberty select cycler, liberty cycler set and the adverse event of sclerotic peritonitis as it was reported this chronic non-infectious condition preexisted pd therapy.The cause of the patient¿s sclerotic peritonitis remains unknown and there was no report pd therapy exacerbated this condition.Sclerotic peritonitis is a rare condition that involves a non-infectious process of peritoneal inflammation of unknown etiology and may be multi-factorial.Due to the confirmation that this patient¿s diagnosis of sclerotic peritonitis predated the start of pd therapy, the liberty select cycler and liberty cycler set can be excluded as the cause of this event.Based on the available information, there is no specific allegation or objective evidence indicating a fresenius device(s) or product(s) deficiency or malfunction, caused or contributed to the patient¿s adverse event.
 
Event Description
It was reported a peritoneal dialysis (pd) patient on continuous cyclic pd (ccpd) therapy on the liberty select cycler is in process of transferring to in-center hemodialysis (hd) due to sclerosing peritonitis.There was no specific allegation this adverse event was due to a deficiency or malfunction of any fresenius product(s) or device(s) in the initial reporting.Upon follow up, the patient¿s peritoneal dialysis registered nurse (pdrn) reported this patient received a diagnosis of sclerotic peritonitis that preexisted pd therapy on an unknown date.It was reported the patient had not experienced an infection and this chronic condition was from unknown etiology.Additionally, the patient has not been hospitalized and did not require additional medical intervention outside ongoing pd therapy support.The patient is currently transitioning to in-center hd therapy as the patient¿s sclerotic peritonitis has increasingly caused complications during ccpd therapy on the liberty select cycler at home.It was confirmed the patient¿s diagnosis of sclerotic peritonitis was not due to a deficiency or malfunction of any fresenius product(s) or device(s).The patient will continue hd for renal replacement therapy on an in-center basis.
 
Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformance's or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE CONN./EXT. DL
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
MDR Report Key10840415
MDR Text Key216322720
Report Number8030665-2020-01672
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861100750
UDI-Public00840861100750
Combination Product (y/n)N
PMA/PMN Number
K043363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number050-87216
Device Catalogue Number050-87216
Was Device Available for Evaluation? No
Device AgeMO
Date Manufacturer Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DELFLEX PD FLUID; LIBERTY SELECT CYCLER
Patient Outcome(s) Required Intervention;
Patient Weight87
-
-