Model Number MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
At this time, the suspect device has not been returned for evaluation.Therefore, no root cause could be determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported that the medisorb¿ multi-absorber canister has a crack at the bottom, which lead to leakage.The event occurred while connected to a patient however, the customer confirmed that no patient harm was associated on this event.
|
|
Manufacturer Narrative
|
The suspect device was not returned for evaluation.However, device history record review reveals no deviations to specification of device.All units packed passed leak test and visual inspection.Batch documentation review reveals no issues that could be linked to the reported failure.Also, devices are packaged in a carton that meets the vyaire drop testing specification and labeled as fragile and to handle with care.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|