• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION HINGE DIS FEM WDG TRL SZ 3 5MM; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION HINGE DIS FEM WDG TRL SZ 3 5MM; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71434229
Device Problems Break (1069); Fracture (1260)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2020
Event Type  malfunction  
Event Description
It was reported that the legion hinge dis fem wdg trial size 3 5mm was found broken.No procedure-related.Therefore, no injuries or surgical delays were reported.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device is fractured, and the fractured piece was returned, rendering the device inoperable.The device was manufactured in 2013 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.Corrected date: d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION HINGE DIS FEM WDG TRL SZ 3 5MM
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10841633
MDR Text Key216306266
Report Number1020279-2020-06455
Device Sequence Number1
Product Code HRY
UDI-Device Identifier00885556032169
UDI-Public00885556032169
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 03/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71434229
Device Catalogue Number71434276
Device Lot Number13EGA0020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2021
Date Manufacturer Received03/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-