• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION AIA-360

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION AIA-360 Back to Search Results
Model Number AIA-360
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2020
Event Type  malfunction  
Manufacturer Narrative
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.Device evaluation by manufacturer: a field service engineer (fse) followed up with the customer over the phone.While troubleshooting with the customer, fse confirmed reported error by reviewing the error logs, and reproduced error by running bf wash, and substrate primes in maintenance.Customer reconfirmed disconnected tubing and reconnected it to the bf probe.The customer successfully performed quality control run without error and results were within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-360 operator's manual under chapter 7, section 7.1: list of error messages states the following: [2015] bf probe purge failure is generated when purging by the bf probe is abnormal.The operator is instructed to clean up the wash probe tip or replace it.Contact the service department.The aia-360 training manual under tab 8: analyzer error codes states the following: [2017] substrate purge failure is generated when no substrate was dispensed at daily maintenance.The operator is instructed to check substrate level and replace as necessary, repeat daily maintenance.The most probable cause of the reported event is due to disconnected bf wash probe tubing.
 
Event Description
A customer reported getting error messages 2015 "bf probe purge failure", and 2017 "substrate purge failure" on the aia-360 analyzer.The customer also reported fluid around the carousel when clearing the quality control cups, and noted tubing to bf probe was found disconnected.Analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for luteinizing hormone (lhii), follicle stimulating hormone (fsh), progesterone (progii), estradiol (e2) and beta human chorionic gonadotropin (bhcg).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
AIA-360
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA   1058623
Manufacturer Contact
bernadette oconnell
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586-23
JA   1058623
MDR Report Key10843720
MDR Text Key216915336
Report Number8031673-2020-00353
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360
Device Catalogue Number019945
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/02/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-