|
Catalog Number UNK_MED |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2020 |
Event Type
malfunction
|
Event Description
|
It was reported via a troubleshooting call that the device had a temperature deviation alarm.The patient was in a rewarming phase on auto mode set to rewarm the patient at 0.5 degrees/hour until the goal temperature of 36.5c was reached.The nurse alleged that patient was rewarming slower than the desired rate.Approximately 6 hours into warming, the axillary temperature read 36.1c, and the esophageal probe read 35.4c.The nurse reported the probe was connected properly and had not been moved.The nurse also stated that no blankets had been removed, and the patient's condition had not changed.The water temperature was 40.1c.
|
|
Manufacturer Narrative
|
The investigation concluded that the cause of the reported event could not be determined, as the customer did not make the device available for evaluation and declined any further troubleshooting.
|
|
Event Description
|
It was reported via a troubleshooting call that the device had a temperature deviation alarm.The patient was in a rewarming phase on auto mode set to rewarm the patient at 0.5 degrees/hour until the goal temperature of 36.5c was reached.The nurse alleged that patient was rewarming slower than the desired rate.Approximately 6 hours into warming, the axillary temperature read 36.1c, and the esophageal probe read 35.4c.The nurse reported the probe was connected properly and had not been moved.The nurse also stated that no blankets had been removed, and the patient's condition had not changed.The water temperature was 40.1c.
|
|
Search Alerts/Recalls
|
|
|