Model Number P3200K000380
Device Problem Self-Activation or Keying (1557)
Patient Problem No Patient Involvement (2645)
Event Date 10/26/2020
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the right caregiver board needed to be replaced.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Head section motor: examine the motor for the presence and tightness of the attachment hardware.Replace as necessary.Fully raise and lower the head section.Make sure there is no friction or abnormal noises and no audible overload indication can be heard during the movement.Replace in the event of a malfunction.Make sure all functions on the caregiver control work correctly.Repair or replace the side rails as necessary.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the right caregiver board to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the head of the bed will raise on its own.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #c-1290804.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
harish vishwanathan
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key10848358
MDR Text Key217696181
Report Number1824206-2020-00481
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP3200K000380
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-