• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY TIBIAL IMPL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY TIBIAL IMPL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, Back to Search Results
Model Number 74018901
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 10/23/2020
Event Type  malfunction  
Event Description
It was reported that during service/pm the journey tibial impl impactor was found to be chipped and worn.No delay and no back up applicable for this event.No injuries involved.
 
Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.Therefore, product analysis could not be performed at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY TIBIAL IMPL IMPACTOR
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10849929
MDR Text Key216560462
Report Number1020279-2020-06519
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010557285
UDI-Public03596010557285
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74018901
Device Catalogue Number74018901
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received04/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-