MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 STR INSRTR THREADED SHAFT; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 STR INSRTR THREADED SHAFT; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Device Damaged Prior to Use (2284); Failure to Advance (2524)
Patient Problem No Information (3190)
Event Date 10/27/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the threads on the instrument are damaged and don't thread into sleeve.No patient involvement.No additional information.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.One g7 str insrtr threaded shaft was returned and evaluated.Upon visual inspection the hex on the threaded end has been stripped.The threads are undamaged.The shaft shows scuffing from use along with the opposite hex.A thread ring gage was able to be fully seated onto the threads of the shaft.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR INSRTR THREADED SHAFT
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10853999
MDR Text Key216689040
Report Number0001825034-2020-04117
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003452
Device Lot Number060059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2020
Was the Report Sent to FDA? No
Date Manufacturer Received03/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-