Model Number N/A |
Device Problems
Device Damaged Prior to Use (2284); Failure to Advance (2524)
|
Patient Problem
No Information (3190)
|
Event Date 10/27/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the threads on the instrument are damaged and don't thread into sleeve.No patient involvement.No additional information.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.One g7 str insrtr threaded shaft was returned and evaluated.Upon visual inspection the hex on the threaded end has been stripped.The threads are undamaged.The shaft shows scuffing from use along with the opposite hex.A thread ring gage was able to be fully seated onto the threads of the shaft.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|