Model Number N/A |
Device Problem
Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/20/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported the inner box was damaged.Another implant was available to use.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned product found the outer carton has been crushed.The inner and outer blister are damaged.Sterility has been compromised.Device history record (dhr) was reviewed and no discrepancies were found.These products were likely conforming when they left zimmer biomet control.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|