Additional suspect medical device components involved in the event: product family: dbs-adapters, upn: m365db9218150, model: db-9218-15, serial: (b)(4), batch: 5180039.The dbs m8 adapters db-9218-15 (serial numbers (b)(4)) were analyzed, passed all tests performed, and exhibited normal device characteristics.The investigation concluded that the reported event was not confirmed through device analysis.Therefore, a probable cause cannot be determined as no problem has been detected.
|
It was reported that the patient underwent a procedure to switch from a non-bsc implantable pulse generator (ipg) to a bsc ipg.During the implant procedure, m8 adaptors were implanted to connect the existing non-bsc leads to the new bsc ipg.An impedance check showed appropriate connectivity so the connection was secured and the patient was sutured.A second impedance check showed high impedances.The physician re-opened the sub clavicular pocket and replaced the m8 adaptors, however, the high impedances remained.The physician decided to keep the system implanted as they could program effective therapy using the normal impedance contacts.
|