Sensor (b)(4) has been returned and investigated.The sensor plug was properly seated, and no physical damage was observed on the sensor patch.Sensor found to be in state 5 (indicating normal termination).No failure modes were observed on the sensor plug assembly upon visual inspection.Sim vivo test was performed, and results were within specification.No malfunction or product deficiency was identified.Dhrs (device history review) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
|