RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 19nov2020.
 
Event Description
It was reported to philips that the device had a battery issue.The device was not being used on a patient at the time of the event.There was no patient or user harm reported.
 
Manufacturer Narrative
G4:18mar2021.B4:20mar2021.The device was evaluated by the customer with assistance from a philips remote service engineer (rse) and it was determined that the battery needed to be replaced to return the device to working condition.The customer¿s bio-med replaced the battery to resolve the issue and bring the device back to functionality.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10868899
MDR Text Key218249232
Report Number2031642-2020-04201
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/27/2020
Initial Date FDA Received11/19/2020
Supplement Dates Manufacturer Received10/27/2020
Supplement Dates FDA Received03/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-