• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN SECTOR W/GRIPTION 52MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN SECTOR W/GRIPTION 52MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-32-052
Device Problem Difficult to Insert (1316)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 11/05/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
During the tightening, the apex hole didn't stop and went beyond the thickness of the cup.The screw was left implanted.
 
Manufacturer Narrative
Product complaint # :(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b7, d4 (lot #) and h4.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.Review of the attached x-ray image could not confirm the reported allegation.The root cause could not be determined.A search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN SECTOR W/GRIPTION 52MM
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10869985
MDR Text Key217177258
Report Number1818910-2020-25037
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295010302
UDI-Public10603295010302
Combination Product (y/n)N
PMA/PMN Number
K071784
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1217-32-052
Device Catalogue Number121732052
Device Lot Number9582402
Was Device Available for Evaluation? No
Date Manufacturer Received02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
APEX HOLE ELIM POSITIVE STOP; APEX HOLE ELIM POSITIVE STOP
Patient Age82 YR
-
-