MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801763
Device Problem Output Problem (3005)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2020
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) could not verify the '2 year service life of your batteries' message.The fsr replaced the batteries.The unit operated to the manufacturer's specifications.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the heart lung machine (hlm) displayed a 'you have exceeded the 2-year service life of your batteries' message earlier than expected.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Updated block: d9 during laboratory analysis, the product surveillance technician (pst) observed both the batteries received with 12.5 volts direct current (vdc) and 375 siemens (s).The batteries were installed into a lab use only (luo) heart lung machine (hlm) and verified that the batteries were charging via the luo central control monitor (ccm).Both batteries were within specification of 52 minutes run time.Per data log review, the system was used on (b)(6) 2020 and gracefully shut down.When the system was powered up next the date had changed to 15-feb-2045.This will trigger the reported message 'you have exceeded the 2-year service life of your batteries'.The next power up the date corrected itself to 26-oct-2020.The log confirms the complaint was caused by an unexpected date change on the ccm.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10874875
MDR Text Key224410227
Report Number1828100-2020-00448
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/20/2020
Supplement Dates Manufacturer Received02/17/2021
03/12/2021
Supplement Dates FDA Received03/12/2021
03/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-