Model Number 801763 |
Device Problem
Output Problem (3005)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/27/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) could not verify the '2 year service life of your batteries' message.The fsr replaced the batteries.The unit operated to the manufacturer's specifications.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the heart lung machine (hlm) displayed a 'you have exceeded the 2-year service life of your batteries' message earlier than expected.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Updated block: d9 during laboratory analysis, the product surveillance technician (pst) observed both the batteries received with 12.5 volts direct current (vdc) and 375 siemens (s).The batteries were installed into a lab use only (luo) heart lung machine (hlm) and verified that the batteries were charging via the luo central control monitor (ccm).Both batteries were within specification of 52 minutes run time.Per data log review, the system was used on (b)(6) 2020 and gracefully shut down.When the system was powered up next the date had changed to 15-feb-2045.This will trigger the reported message 'you have exceeded the 2-year service life of your batteries'.The next power up the date corrected itself to 26-oct-2020.The log confirms the complaint was caused by an unexpected date change on the ccm.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|