SMITH & NEPHEW, INC. GII NON POR FIN PUNCH 3-4; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71440266 |
Device Problem
Dull, Blunt (2407)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that, after a tka surgery was performed, it was found that the fin punch is old and dull.The procedure had been completed with that instrument.The patient was not involved at the time of finding.
|
|
Manufacturer Narrative
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is dull and has several nicks and scratches.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|