Catalog Number V041163000000
Device Problems Electrical /Electronic Property Problem (1198); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where stalling has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a silver reciproc caused a short circuit as the assistant connected the device with the power supply.The fuse blows out.No injury occured.
 
Manufacturer Narrative
Battery (voltage breaks down under load) defective, replaced.Charging socket (loose contact) defective, replaced.Micromotor smr116048, contra-angle handpiece 26929 (2011) and foot control 58113 are not defective.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW SILVER RECIPROCATING MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM  81737
MDR Report Key10881522
MDR Text Key218937480
Report Number9611053-2020-00351
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 01/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberV041163000000
Date Manufacturer Received01/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-