This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.More then 5 years have passed since the subject device was manufactured.Based on the results of the investigation, the likely causes are as follows: missing glue and cut on probe unit: it is likely due to aging deterioration or due to an external force (such as strong rubbing against the relevant part during normal use, including reprocessing).Forceps elevator is not sitting properly: it is likely an excessive force applied caused an extension of the wires that perform the forceps rise-up table manipulation.The instructions for use include the following which can detect the event: "inspection of the endoscope: inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.".Per the legal manufacturer, the other device defects included in the initial mdr have no potential to cause or contribute to death or serious injury if the malfunctions were to recur.
|