• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN PROVISIONAL STEM; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN PROVISIONAL STEM; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problem Foreign Body In Patient (2687)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Event date: unknown day in september, 2020.Implant date: unknown day in september, 2020.Explant date: unknown day in september, 2020.Customer has indicated that the product will not be returned to zimmer biomet for investigation, due to hospital policy.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2020 - 03877.
 
Event Description
It was reported that during right total hip arthroplasty the plastic trial head dislodged from the trail neck/stem while reducing the hip.The surgeon made multiple attempts to retrieve the trail head but was unsuccessful.The procedure was completed and the patient sent to recovery.While in recovery, a trauma surgeon was consulted and the patient was brought back to surgery the same day and the retained item was removed.The patient tolerated the procedures well.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
No product was returned; visual and dimensional evaluations could not be performed.Medical records were not provided.Part and lot identification are necessary for review of device history records, neither were provided for the stem provisional.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN PROVISIONAL STEM
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10884691
MDR Text Key217672473
Report Number0001822565-2020-03876
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-