Model Number N/A |
Device Problem
Failure to Cut (2587)
|
Patient Problems
Abrasion (1689); Scarring (2061)
|
Event Date 11/06/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that a dermatome malfunction occurred in which it was not cutting properly.The event occurred during surgery.According to dr.Van brussel the patient experienced a full thickness cut, scar wounds, and had to be stitched.No additional consequences have been reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the control bar was out of position and the unit was out of calibration at the 0, 4, and 12 readings.The vespel and semi-circle bearings were replaced to tighten the torque on the thickness control lever.Additionally, the control bar was placed in the correct position and the device was calibrated to resolve the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|
Event Description
|
There is no additional information.
|
|