Model Number 09H39-01
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/03/2020
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
 
Event Description
The customer reported a hemoglobin mismatch occurring on the cell-dyn emerald analyzer on an (b)(6) yr old male patient.Results provided: (b)(6) 2020 sid (b)(6) = 9.2 g/dl (reference range: 11-18 g/dl), another lab = 6.9 g/dl (reference range: 14-18 g/dl).All available patient information was included.Additional patient details are not available.No impact to patient management was reported.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs and a review of labeling.No trends were identified for the customers issue.Labeling was reviewed and found to be adequate.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN EMERALD SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
ABBOTT LABORATORIES
4551 great america parkway
santa clara CA 95054
MDR Report Key10886667
MDR Text Key223393807
Report Number2919069-2020-00036
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740020088
UDI-Public00380740020088
Combination Product (y/n)N
PMA/PMN Number
K081495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number09H39-01
Device Catalogue Number09H39-01
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age88 YR
-
-