Model Number AS480
Device Problem Failure to Pump (1502)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2020
Event Type  malfunction  
Manufacturer Narrative
Root cause: potential alteration of staining in this case was due to improper operation of the buffer pump.The problem was solved by field service engineer with replacement of the part.Following the replacement, the instrument was fully operational within specifications, without errors and available for the user.Failure mode description: following a buffer pump malfunction or if the part stops working; the resulting failure modes described could occur: dripping from the probe and reagents to remain to the slides.Failure mode in all scenarios has the potential to cause inconsistent staining.
 
Event Description
Customer complaint record reported the event as follows: dripping from the probe.No direct or indirect patient harm or user harm have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSTAINER LINK 48
Type of Device
AUTOSTAINER LINK 48
Manufacturer (Section D)
DAKO NORTH AMERICA, INC
6392 via real
carpinteria CA 93013
Manufacturer (Section G)
DAKO NORTH AMERICA, INC.
6392 via real
carpinteria CA 93013
Manufacturer Contact
sonia siddique
6392 via real
carpinteria, CA 93013
5123328112
MDR Report Key10896475
MDR Text Key220964219
Report Number2022180-2020-00391
Device Sequence Number1
Product Code KPA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAS480
Device Catalogue NumberAS48030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2010
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-