Model Number GF-UE160-AL5
Device Problem Poor Quality Image (1408)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.During the evaluation, the pink rubber was found with a deep cut.The image was visible after aeration and the ultrasound image contained one broken element.All the camera switches and the scope connectors were working properly after aeration.The control knob was found with low tensions.No other issues were found during the device inspection.If additional information is obtained a supplemental report will be filed.
 
Event Description
A user facility reported image issues on a gastrovideoscope.No patient involvement or injury was reported.No additional information has been obtained.
 
Manufacturer Narrative
The investigation has been completed.A device history record (dhr) review was performed and found no non-conformances that would cause the reported malfunction.All records indicate that the device was manufactured in accordance with all applicable procedures.The instructions for use (ifu) states: ¿do not apply shock to the distal end of the insertion tube, particularly the ultrasonic transducer and the objective lens surface at the distal end.¿ the root cause could not be conclusively determined.Probable causes that could have led to the reported event include that the water immersion in the interior may have led to the occurrence of this phenomenon.In addition of an external force might have been applied to the distal end of the acoustic lens because the image showed that the acoustic lens was damaged, and there was also a lack of sound line.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC GASTROVIDEOSCOPE
Type of Device
ULTRASONIC GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10900770
MDR Text Key218224259
Report Number8010047-2020-09423
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
PMA/PMN Number
K051541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UE160-AL5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Was the Report Sent to FDA? No
Date Manufacturer Received01/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-