Model Number 1192 |
Device Problem
Installation-Related Problem (2965)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/13/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results, method, and conclusion codes along with the investigation results will be provided in final report.
|
|
Event Description
|
Related manufacturer reference number: 3006705815-2020-32886.It was reported that during an elective ipg procedure the patients lead migrated out of the epidural space.Upon inspection it was found that the lead anchor had opened resulting in the migration.During the procedure the lead and anchor were explanted and replaced.
|
|
Manufacturer Narrative
|
The reported event of functional issue of the anchor was not confirmed.As received, the anchor was post capa 140614 and complete, and worked as design.No root cause was identified for the reported event.
|
|
Search Alerts/Recalls
|